INRATIO PT/INR TEST STRIPS
Report
- Report Number
- 2027969-2014-00745
- Event Type
- Malfunction
- Date Received
- August 12, 2014
- Date of Event
- July 30, 2014
- Report Date
- July 30, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
INVESTIGATION/CONCLUSION: IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVALUATION. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, A REVIEW OF IN-HOUSE TESTING DATA WAS PERFORMED. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCTS PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. ALTHOUGH THE ROOT CAUSE ANALYSIS DID NOT INCLUDE RETURN TESTING, IMPROPER TECHNIQUES WERE IDENTIFIED IN THE COMPLAINT. THESE COULD NOT BE RULED OUT AS A CAUSE OF THE UNEXPECTED RESULTS. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED. THE LOT MET SPECIFICATIONS AND NO NON-CONFORMANCES WERE DOCUMENTED. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT INRATIO2 INR RESULTS. RESULTS ARE AS FOLLOWS: DATE: (B)(6) 2014; INRATIO2 INR: 5.2 AND 1.9. THE TIME BETWEEN TESTING WAS 1.5 HOURS. REPORTEDLY, MILKING FINGER AFTER FINGER STICK, UNABLE TO OBTAIN SUFFICIENT BLOOD SAMPLE. THERAPEUTIC RANGE 2.0 - 3.0 FOR THE PATIENT. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 478975 | INRATIO PT/INR TEST STRIPS | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 344682 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | COUMADIN| UNSPECIFIED PAIN MEDICATIONS| PATIENT RECENTLY DISCONTINUED ULTRAM| INRATIO MONITOR: SN (B)(4) |