FDA Adverse Event Injury Summary report: N

ZILVER PTX DRUG-ELUTING PERIPHERAL STENT

MDR report key: 4071558 · Received August 28, 2014

Report

Report Number
3001845648-2014-00156
Event Type
Injury
Date Received
August 28, 2014
Date of Event
September 17, 2013
Report Date
July 30, 2014
Manufacturer
COOK IRELAND LTD
Product Code
NIU
PMA / PMN Number
P100022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL PMA/510(K) NUMBER: S001. THIS REPORT IS BEING SUBMITTED FOR THE DISTAL ZILVER PTX DEVICE REPORTED FOR STRIKING ANEURYSMAL SPACES/PSEUDOANEURYSM (ZIV6-35-125-6.0-120-PTX). THE DEVICE INVOLVED IN THIS COMPLAINT WAS IMPLANTED IN THE PATIENT AND THEREFORE NOT AVAILABLE TO BE RETURNED FOR EVALUATION; WITH THE INFORMATION PROVIDED A DOCUMENT BASED INVESTIGATION WAS CARRIED OUT. THE DEVICE INVOLVED IN THIS COMPLAINT IS: ZIV6-35-125-6.0-120-PTX OF LOT NUMBER C781685 OR C780023. THE SPECIFIC LOT NUMBER COULD NOT BE DISTINGUISHED AS BOTH STENTS ARE IDENTICAL ON IMAGING AND THE COMPLAINT REPORT PROVIDES NO INFORMATION ON THE SPECIFIC LOCATION OF EACH 6MM STENT. A REVIEW OF THE RELEVANT MANUFACTURING RECORDS DID NOT REVEAL ANY DISCREPANCY WHICH COULD HAVE CONTRIBUTED TO THE COMPLAINT ISSUE. ANGIOGRAM IMAGES WERE PROVIDED TO SUPPORT THE COMPLAINT INVESTIGATION AND THE FOLLOWING CLINICAL IMAGE REVIEW WAS PROVIDED: "FINDINGS: THE INITIAL ANGIOGRAPHY DEMONSTRATES PRIMARY STENTING OF A LONG LEFT SUPERFICIAL FEMORAL ARTERY (SFA) OCCLUSION FROM FEMORAL AND POPLITEAL ACCESS POINTS. THE ENTIRE SFA WAS OCCLUDED. RECANALIZATION WAS PERFORMED FROM DISTAL TO PROXIMAL USING SUB-INTIMAL TECHNIQUE; HOWEVER, THE WIRE WAS MUCH FARTHER OUTSIDE THE LUMEN THAN USUAL AND AT LEAST IN THE ADVENTITIA IF NOT EXTRAVASCULAR. REVASCULARIZATION WAS ACCOMPLISHED WITH THREE OVERLAPPING ZILVER PTX STENTS EXTENDING FROM THE MID LEFT COMMON FEMORAL ARTERY (CFA) TO THE DISTAL LEFT SFA PAST THE ADDUCTOR CANAL. A LEFT COMMON FEMORAL ARTERY ANEURYSM WAS PRESENT AND THE PATIENT WAS REPORTED TO HAVE AN AAA BUT NO IMAGING OF THIS WAS PROVIDED. FOLLOW-UP LEFT LEG ANGIOGRAPHY APPARENTLY PERFORMED COINCIDENTALLY AT THE TIME OF CORONARY ANGIOGRAPHY WAS PERFORMED 11 AND 12 MONTHS POST INITIAL STENTING. ANGIOGRAPHY AND STENTING OF THE RIGHT CORONARY ARTERY WAS PERFORMED (B)(6) 2013. NO FOLLOW UP IMAGING OF AN UNCOATED CORONARY STENT AS SUGGESTED IN THE COMPLAINT REPORT IS PROVIDED. MULTIPLE STENT FRACTURES ARE EVIDENT ON EACH FOLLOW UP ANGIOGRAMS INVOLVING THE PROXIMAL AND DISTAL MOST ZILVER STENTS. STRIKING ANEURYSMAL SPACES HAVE DEVELOPED IN THE SFA WALL AND THE STENTS ON ANGIOGRAPHY. THE SPACES ARE LARGER PROXIMALLY AND LESSEN DISTALLY ONCE THE MID SFA STENT IS REACHED. THE SPACE AT THE PROXIMAL DISCONNECTION IS SIMILAR TO THE ADJACENT SPACES AND THE ARTERY IS NEAR NORMAL AT THE MID AND DISTAL STENT OVERLAP POINT. THE ANGIOGRAM DID NOT APPRECIABLY CHANGE DURING THE MONTH BETWEEN FOLLOW-UPS. THE ANEURYSMAL SPACES ARE CONSISTENT WITH PSEUDOANEURYSMS. THE PSEUDOANEURYSMS ARE LARGEST AROUND THE MIDDLE AND PROXIMAL STENTS. POST RECANALIZATION AND ANGIOPLASTY ANGIOGRAPHY DEMONSTRATED A DISSECTED AND OCCLUDED SFA MATCHING THE FUTURE PSEUDOANEURYSM LOCATION. ADDITIONALLY THE SEVERITY OF THE PSEUDOANEURYSM FORMATION MATCHES THE DEGREE TO WHICH THE RECANALIZATION TRACT STRAYED FROM THE VESSEL TRUE LUMEN. PROXIMALLY WHERE THE TRUE LUMEN WAS AT LEAST 4-5 MM OUTSIDE THE TRUE LUMEN, THE PSEUDOANEURYSMS WERE MOST SEVERE. DISTALLY WHERE THE CHANNEL WAS CLOSER TO TRUE LUMEN, THE PSEUDOANEURYSMS WERE LESS SEVERE. IMPRESSION: STRIKING SPACES DID DEVELOP BETWEEN THE STENTS AND THE SFA WALL PARTICULARLY IN THE PROXIMAL AND MID LEFT SFA. THE MOST LIKELY EXPLANATION IS STENT PLACEMENT OUTSIDE (IN THE ADVENTITIA) OF THE ARTERY WITH THE ANEURYSMAL APPEARANCE REPRESENTING PSEUDOANEURYSM FORMATION. CONCLUDING THAT THIS WAS IMPOSSIBLE GIVEN THE NORMAL IMMEDIATE POST STENT APPEARANCE MIGHT BE CONSIDERED REASONABLE AT FACE VALUE; HOWEVER, A CASE DEMONSTRATING A NORMAL IMMEDIATE POST STENT APPEARANCE WITH DELAYED PSEUDOANEURYSM FORMATION HAS BEEN PUBLISHED (J VASC SURG 2010; 52:216-8). INFECTION IS POSSIBLE BUT UNLIKELY GIVEN THE LACK OF CLINICAL DATA SUPPORTING INFECTION. RARELY UNRECOGNIZED CONNECTIVE TISSUE DISORDER CAN RESULT IN ANEURYSM FORMATION AFTER CHARACTERIZATION; HOWEVER, THESE CASES USUALLY OCCUR IN YOUNG PATIENTS. GIVEN DRUG COATING MUST CREATE DILATED PERIVASCULAR SPACES AT A CELLULAR LEVEL, PERIVASCULAR SPACE SIZE SHOULD NOT DEPEND ON THE ORIGINAL VESSEL DIAMETER. ALTHOUGH SIZES OF DILATED PERIVASCULAR SPACES ARE NOT WELL DESCRIBED, THE CORONARY LITERATURE PROVIDES A GENERAL RANGE BETWEEN 0.7 MM TO 1.7 MM (CIRCULATION. 2011;123:2382-2391) WITH LARGER SPACES CALLED ANEURYSMS. THE PERIVASCULAR SPACES IN THIS CASE ARE MUCH LARGER THAN WOULD BE ANTICIPATED FROM LOCAL DRUG EFFECT AND ALSO SHOULD BE CALLED ANEURYSMS. ADDITIONALLY, THESE SPACES ARE EASILY RECOGNIZABLE AND IF SECONDARY TO PACLITAXEL SHOULD HAVE BEEN EVIDENT NOT ONLY IN THE CLINICAL TRIAL BUT IN SUBSEQUENT PEER REVIEW LITERATURE. HOWEVER, ANEURYSMS HAVE YET TO BE REPORTED IN THE SFA. THE STENT FRACTURES ARE SECONDARY TO INCREASED MOBILITY OF THE STENTS INSIDE THE ANEURYSMAL SFA. SFA THROMBOSIS SECONDARY TO STENT FRACTURE IS POSSIBLE." AS THE DEVICE WAS NOT AVAILABLE FOR EVALUATION AND CONDITIONS OF USE CANNOT BE REPLICATED IN THE LABORATORY, IT IS NOT POSSIBLE TO CONCLUSIVELY DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. PRIOR TO DISTRIBUTION ALL ZILVER PTX DEVICES ARE SUBJECT TO VISUAL INSPECTION AND FUNCTIONAL CHECKS TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE RELEVANT MANUFACTURING RECORDS FOR ZILVER PTX AND ZILVER PTX DRUG ELUTING STENT REVEALED NO DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT. AS PER THE DEVICES' INSTRUCTIONS FOR USE, ARTERIAL ANEURYSM AND PSEUDOANEURYSM FORMATION ARE NOTED AS POTENTIAL ADVERSE EVENTS ASSOCIATED WITH THE PLACEMENT OF THESE DEVICES. THE PATIENT CONTINUES TO BE MONITORED AND NO FURTHER ADVERSE EFFECTS HAVE BEEN REPORTED. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL MERGING TRENDS.

Description of Event or Problem · 1

ON (B)(6) 2012, THREE PTX STENTS (1X ZIV6-35-125-7.0-120-PTX AND 2X ZIV6-35-125-6.0-120-PTX) WERE IMPLANTED IN THE LEFT SFA OF A PATIENT. THE BLOOD FLOW OF THE LESION AFTER THE PROCEDURE WAS GOOD. FOLLOW UP ANGIOGRAPHY WAS CONDUCTED ONE YEAR AFTER INITIAL PROCEDURE. A CLINICAL REVIEW OF IMAGES PROVIDED IN RELATION TO THIS EVENT REVEALED MULTIPLE STENT FRACTURES ON THE PROXIMAL AND MOST DISTAL ZILVER STENTS. THE DISTAL STENT WAS TRANSECTED AND THE DISTAL SEGMENT DISPLACED INFERIORLY. THE PROXIMAL STENT HAD TWO SITES OF FRACTURE AND HAD ALSO MIGRATED SUPERIORLY. STRIKING ANEURYSMAL SPACES WERE ALSO NOTED TO HAVE DEVELOPED IN THE SFA WALL AND THE STENTS. THE SPACES ARE LARGER PROXIMALLY AND LESSEN DISTALLY ONCE THE MID SFA IS REACHED. AS THE BLOOD FLOW WAS GOOD, NO INTERVENTION HAS BEEN PERFORMED TO DATE. THE PHYSICIAN PLANS TO PERFORM A FOLLOW UP ACTION. AS PER THE COMPLAINT DESCRIPTION RECEIVED, THREE DEVICES ARE INVOLVED IN THIS INCIDENT. ADDITIONAL SEPARATE REPORTS WILL BE SUBMITTED IN RELATION TO THE OTHER DEVICE REPORTED - REPORT REFERENCE NUMBERS: 3001845648-2014-00154 AND 3001845648-2014-00155.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
525052 ZILVER PTX DRUG-ELUTING PERIPHERAL STENT NIU STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING NIU COOK IRELAND LTD C781685 OR C780023

Patients

Seq Age Sex Outcome Treatment
1 66 YR Life Threatening