NOMAD PRO
MDR report key: 4071212
·
Received September 2, 2014
Report
- Report Number
- 3005417494-2014-00001
- Event Type
- Malfunction
- Date Received
- September 2, 2014
- Date of Event
- July 24, 2014
- Report Date
- August 30, 2014
- Manufacturer
- ARIBEX, INC.
- Product Code
- EHD
- PMA / PMN Number
- K935046
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE HANDSET AND THE DEVICE WERE ASSEMBLED ON 17-OCT-2011 AND IS ABOUT 3 YEARS OLD. EXPECTED END OF LIFE FOR THE HANDSET IS 2 YEARS OR 300 CHARGE/DISCHARGE CYCLES. THE BATTERY PACK (HANDSET) WAS BEYOND END OF LIFE SINCE THE BATTERY WAS OVER 2 YEARS OLD. ADDITIONAL INVESTIGATION IS ONGOING AND A FOLLOW-UP REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE BATTERY PACK (HANDSET) OF THE DEVICE CAUGHT FIRE WHILE PLACED IN A DRAWER AND PARTIALLY MELTED.