FDA Adverse Event Malfunction Summary report: N

NOMAD PRO

MDR report key: 4071212 · Received September 2, 2014

Report

Report Number
3005417494-2014-00001
Event Type
Malfunction
Date Received
September 2, 2014
Date of Event
July 24, 2014
Report Date
August 30, 2014
Manufacturer
ARIBEX, INC.
Product Code
EHD
PMA / PMN Number
K935046
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE HANDSET AND THE DEVICE WERE ASSEMBLED ON 17-OCT-2011 AND IS ABOUT 3 YEARS OLD. EXPECTED END OF LIFE FOR THE HANDSET IS 2 YEARS OR 300 CHARGE/DISCHARGE CYCLES. THE BATTERY PACK (HANDSET) WAS BEYOND END OF LIFE SINCE THE BATTERY WAS OVER 2 YEARS OLD. ADDITIONAL INVESTIGATION IS ONGOING AND A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE BATTERY PACK (HANDSET) OF THE DEVICE CAUGHT FIRE WHILE PLACED IN A DRAWER AND PARTIALLY MELTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
532451 NOMAD PRO EXTRAORAL SOURCE X-RAY SYSTEM EHD ARIBEX, INC. FP-0005

Patients

Seq Age Sex Outcome Treatment
1