FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4070895
·
Received September 9, 2014
Report
- Report Number
- 1031452-2014-08624
- Event Type
- Malfunction
- Date Received
- September 9, 2014
- Report Date
- August 12, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER INDEPENDENT REPAIR CENTER STATEMENT THE UNIT HAD LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS THE BARB FITTING FOR THE SIEVE BEDS WAS LEAKING. ADDITIONAL MALFUNCTIONS INCLUDE THE TIE WRAPS FOR THE SIEVE BEDS WERE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 555999 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |