FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4070885
·
Received September 9, 2014
Report
- Report Number
- 1031452-2014-08611
- Event Type
- Malfunction
- Date Received
- September 9, 2014
- Report Date
- August 12, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER INDEPENDENT REPAIR CENTER STATEMENT THE UNIT HAS LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS THE PE VALVE FOR THE SIEVE BEDS WAS DEFECTIVE. ADDITIONAL MALFUNCTIONS IN THE ZIP TIES FOR THE SIEVE BEDS NEEDED TO BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 555624 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC10LX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |