FDA Adverse Event Injury Summary report: N

MAESTRO MINI-TWIN BRACKET WITH HOOK

MDR report key: 4068001 · Received August 12, 2014

Report

Report Number
2081322-2014-00001
Event Type
Injury
Date Received
August 12, 2014
Date of Event
July 15, 2014
Report Date
August 12, 2014
Manufacturer
ORTHO ORGANIZERS, INC.
Product Code
EJF
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TOOTH FRACTURE CAN OCCUR UPON DEBONDING (REMOVAL) OF BRACKET. ALTHOUGH MOST FRACTURES ARE LIMITED TO ENAMEL, THE POTENTIAL FOR PULPAL INVOLVEMENT EXISTS. LITERATURE DESCRIBING DEBONDING ADVERSE EVENTS CAN BE TRACKED BACK TO AT LEAST 1996. PROVIDERS ARE TRAINED TO USE CARE WHEN DEBONDING BRACKETS.

Description of Event or Problem · 1

ON (B)(6) 2013, BRACKETS WERE PLACED. INITIAL DIAGNOSIS: CLASS I 18 MONTHS, MODERATE CROWDING. TOOTH #11 PERFECTLY HEALTHY WITH NO SIGNS OF PREVIOUS RESTORATIONS, DECAY, OR DECALCIFICATION. DENTIST WAS REMOVING BRACKET FROM UL3 (TOOTH #11) AND TOOTH BROKE HORIZONTALLY IN HALF. POST INCIDENT DIAGNOSIS: #11 HORIZONTAL FRACTURE WITH SMALL PULPAL INVOLVEMENT. NO SIGNS OF DECALCIFICATION OR DECAY. CLEAN FRACTURE WITH BRACKET STILL ATTACHED TO THE FRACTURED PORTION. 2 LUGS BROKEN DURING REMOVAL. TOOTH #11 WILL NEED ENDO WITH BUILD UP AND EMAX CROWN AT A LATER DATE. TOOTH WAS BONDED BACK TOGETHER AND NEW BRACKET PLACED ON TOOTH. (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
479095 MAESTRO MINI-TWIN BRACKET WITH HOOK EJF ORTHO ORGANIZERS, INC. 711-399 UNK

Patients

Seq Age Sex Outcome Treatment
1 17 YR Required Intervention TRIUMPH BRACKET REMOVER, PN 205-220XL, PLIERS| TRANSBOND XT, PN 133-0606, BRACKET ADHESIVE