FDA Adverse Event Malfunction Summary report: N

V-60 VENTILATOR

MDR report key: 4066409 · Received August 29, 2014

Report

Report Number
4066409
Event Type
Malfunction
Date Received
August 29, 2014
Date of Event
August 11, 2014
Report Date
August 29, 2014
Manufacturer
RESPIRONICS CALIFORNIA INC.
Product Code
CBK
Product Problem
Yes
Report Source
User Facility report
Reporter Location
OH, US

Narratives

Description of Event or Problem · 1

SYMPTOM: THE ORIGINAL DESIGN NAV-RING FUNCTIONALITY IS UNRESPONSIVE OR INTERMITTENT.CAUSE: THE ORIGINAL DESIGN FRONT BEZEL HAS A LOCATING NOTCH (FIGURE 1) MOLDED INTO THE PLASTIC AND THE NAV-RING OVERLAY HAS A LOCATING CUT-OUT, BOTH WERE USED FOR ALIGNMENT DURING MANUFACTURING AND SERVICE. IT HAS BEEN DETERMINED THAT THE NAV-RING CUT-OUT IS SUSCEPTIBLE TO LIQUID INGRESS CAUSING THE NAV-RING TO BE NON-RESPONSIVE. THIS TYPICALLY OCCURS WHEN CLEANING AGENTS WERE SPRAYED DIRECTLY AND REPEATEDLY ONTO THE NAV-RING ASSEMBLY.======================MANUFACTURER RESPONSE FOR RESPIRONICS V-60 VENTILATOR, V-60 VENTILATOR (PER SITE REPORTER).======================ACTION FOR PERFORMANCE - RETROFIT ON FAILURE.REMEDY: REPLACE THE ORIGINAL NAV-RING BY INSTALLING A 2ND GENERATION FRONT BEZEL WHICH REMOVE THE MOLDED NOTCH IN THE UI FRONT BEZEL AS WELL AS THE CUT-OUT IN THE NAV-RING OVERLAY. THIS WILL IMPROVE THE ROBUSTNESS OF THE NAV-RING ASSEMBLY AND GREATLY REDUCE THE SUSCEPTIBILITY OF LIQUID INGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
527982 V-60 VENTILATOR VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS CALIFORNIA INC. V-60 *

Patients

Seq Age Sex Outcome Treatment
1 *