ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-25233
- Event Type
- Injury
- Date Received
- September 5, 2014
- Report Date
- August 29, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN HAS A PINKISH CONTRAST. THE TOTAL DAILY DOSE ADDS UP CORRECTLY AND REFLECTS THE USERS PROGRAMMED BASAL RATE. TYPICAL USAGE OBSERVED IN ALARM HISTORY. THE PUMP SUCCESSFULLY PASSED A DELIVERY ACCURACY TEST AND WAS FOUND TO BE OPERATING WITHIN REQUIRED SPECIFICATIONS AND DELIVERING ACCURATELY. BATTERY CAP AND CARTRIDGE CAP WERE NOT RETURNED; TEST CAPS USED TO COMPLETE TESTING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER STATED THAT THE PATIENT COULD NOT READ THE SCREEN SAFELY. THE REPORTER STATED THAT THE PATIENT HAD A BLOOD GLUCOSE (BG) OF 30MMOL/L WITH POLYDIPSIA, POLYURIA, AND SYMPTOMS OF DEHYDRATION. THE PATIENT DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE CARE OF DIABETES MANAGEMENT. THE REPORTER STATED THAT BOLUS SETTING ADJUSTMENTS MADE. THIS REPORT IS BEING MADE DUE TO BG EXCURSION THE PATIENT EXPERIENCED DUE TO AN ALLEGED DISPLAY ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 546759 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Life Threatening |