ANIMAS VIBE
Report
- Report Number
- 2531779-2014-25209
- Event Type
- Malfunction
- Date Received
- September 5, 2014
- Report Date
- August 19, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/17/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED AT THE OPENING. A TEST BATTERY CAP WAS USED TO COMPLETE ALL TESTING. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE DISPLAY WAS DIM AND FADED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THE DISTRIBUTOR ALLEGED THAT THE BATTERY COMPARTMENT WAS CRACKED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 544494 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |