FDA Adverse Event
Malfunction
Summary report: N
MISTRAL AIR-PLUS
MDR report key: 4062551
·
Received August 18, 2014
Report
- Report Number
- 4062551
- Event Type
- Malfunction
- Date Received
- August 18, 2014
- Date of Event
- August 14, 2014
- Report Date
- August 18, 2014
- Manufacturer
- STRYKER MEDICAL
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MD, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
MASKING TAPE FOUND ON PATIENT WARMER WRAPPED AROUND HOSING. TAPE CAUGHT FIRE AND WAS IMMEDIATELY PUT OUT. NO HARM TO PATIENT OR PERSONNEL. WARMER REMOVED FROM SERVICE.FROM BIO-MED: STRYKER MODEL MA1100-PM (MISTRAL AIR) IS THE DEVICE WE NOTED THAT NEARLY STARTED A FIRE. WHAT APPEARED TO HAPPEN WAS THERE HAD BEEN A SPLIT IN THE HEATER HOSE, WHICH STAFF PLACED SOME TAPE ON IT TO SEAL THE AIR. THAT TAPE ACTED AS THE COMBUSTIBLE MATERIAL WHICH THEN CAUSED THE PURPLE PLASTIC TO MELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 496391 | MISTRAL AIR-PLUS | SYSTEM, THERMAL REGULATING | DWJ | STRYKER MEDICAL | MA1100-PM | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | NO. |