FDA Adverse Event Malfunction Summary report: N

ROTATABLE SNARE

MDR report key: 4062540 · Received September 5, 2014

Report

Report Number
3005099803-2014-02917
Event Type
Malfunction
Date Received
September 5, 2014
Date of Event
January 14, 2014
Report Date
August 12, 2014
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Product Code
FDI
PMA / PMN Number
K992477
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS FINDING THE FLARE DETACHED. INVESTIGATION RESULTS: VISUAL EVALUATION OF THE RETURNED DEVICE FOUND THAT THE FLARE WAS DETACHED AND THE CATHETER WAS SLIGHTLY KINKED. EVIDENCE OF THE FLARE MANUFACTURING PROCESS WAS PRESENT ON THE CATHETER. FUNCTIONAL TESTING COULD NOT BE PERFORMED DUE TO THE FLARE DETACHMENT. THE COMPLAINT WAS CONFIRMED; THE FLARE DETACHMENT PREVENTED THE SNARE LOOP FROM EXTENDING. HOWEVER, THE FLARE DETACHED IS CONTRARY TO WHAT WAS ORIGINALLY REPORTED. THE REVIEW AND ANALYSIS OF ALL AVAILABLE INFORMATION FAILED TO INDICATE THE ROOT CAUSE OF THIS COMPLAINT. THERE IS AN INVESTIGATION IN PLACE TO ADDRESS THIS ISSUE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION. A SEARCH OF THE COMPLAINT DATABASE CONFIRMED THAT NO OTHER COMPLAINTS EXIST FOR THE SPECIFIED BATCH.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ROTATABLE MEDIUM OVAL SNARE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE LOOP OF THE SNARE DID NOT COME OUT OF THE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER ROTATABLE SNARE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THE FLARE DETACHED, THEREFORE THIS IS NOW AN MDR REPORTABLE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
544825 ROTATABLE SNARE SNARE, FLEXIBLE FDI BOSTON SCIENTIFIC - COSTA RICA (COYOL) M00561830 16033264

Patients

Seq Age Sex Outcome Treatment
1