EXTRACTOR ASNIS III FOR SCREWS:Ø4.0MM AO FITTING
Report
- Report Number
- 0008031020-2014-00414
- Event Type
- Malfunction
- Date Received
- September 5, 2014
- Date of Event
- August 8, 2014
- Report Date
- August 8, 2014
- Manufacturer
- STRYKER GMBH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
THE REPORTED INCIDENT THAT EXTRACTOR ASNIS III FOR SCREWS:Ø4.0MM AO FITTING WAS ALLEGED OF ISSUE S-18 (INSTRUMENT BREAKAGE IDENTIFIED OUT OF SURGERY) COULD BE CONFIRMED. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO A USER RELATED ISSUE. THE FAILURE WAS CAUSED BY INADEQUATE MAINTENANCE OF THE DEVICE. THE DEVICE INSPECTION REVEALED THE FOLLOWING: TIP BREAKAGE COULD BE CONFIRMED. BROADLY SPEAKING, THE DEVICE APPEARED IN GOOD CONDITION, BUT SOME RUST WAS IDENTIFIED ON THE BREAKAGE SURFACE. PLEASE NOTE THAT THE CLEANING AND STERILIZATION GUIDE (L24002000-EN REV. M, OT-RG-1 REV. 1 CLEANING AND STERILIZATION GUIDE) REPORTS THE FOLLOWING: ''THE GUIDANCE CONCENTRATIONS AND TIMES FOR DEVICE IMMERSION IN THE CLEANING SOLUTIONS AND/OR DISINFECTANTS GIVEN BY THE DETERGENT MANUFACTURERS SHALL BE OBSERVED. IF THESE CONCENTRATIONS AND TIMES ARE EXCEEDED SIGNIFICANTLY, DISCOLORATION OR CORROSION COULD OCCUR WITH SOME MATERIALS. THIS COULD ALSO HAPPEN IF RINSING AFTER CLEANING AND/OR DISINFECTING IS INSUFFICIENT." A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. GIVEN THE HIGH FREQUENCY OF THIS FAILURE MODE, THIS CASE HAS BEEN PRESENTED TO R&D DEPARTMENT, WHICH ANALYSES THE EVENT AND PROVIDED THE FOLLOWING STATEMENT "A SUCCESSFUL USAGE OF THE EXTRACTOR CANNOT BE GUARANTEED NOT HAS IT BEEN VALIDATED BECAUSE THE IMPLANT TO BE REMOVED DOES NO LONGER REFLECT THE IMPLANT DESIGN OF THE SALABLE ITEM. ADDITIONAL FACTORS BESIDES THE MISSING GEOMETRICAL FITTING INFLUENCING THE REMOVAL CHANCE WITH THE EXTRACTOR DEVICE ARE ALSO THE BONE HARDNESS, GRADE OF BONY INGROWN, ACCESSIBILITY OF THE IMPLANT, SURROUNDING LIGAMENTS, MUSCLES, TENDONS, SOFT TISSUE, ETC. THESE FACTORS ARE DIFFERENT FROM CASE TO CASE AND SOMETIMES DIFFICULT TO ASSESS (BONE QUALITY) AT THE TIME OF THE EMERGENCY REMOVAL. TO CONCLUDE: A FAILURE OR EVEN DAMAGE OF THE EMERGENCY INSTRUMENTS IS TAKEN INTO CONSIDERATION AS IT IS AT THIS STAGE THE LAST RESORT FOR AN IMPLANT REMOVAL WITHOUT FURTHER DESTRUCTION OF THE IMPLANT ITSELF OR RESECTION OF THE SURROUNDING BONE AND SOFT TISSUE. A HIGHER BREAKAGE RATE OF EXTRACTION INSTRUMENTS AND THUS THE ASNIS III EXTRACTOR IS ACCEPTABLE. NO DESIGN CHANGE IS NECESSARY AT THIS TIME." IF ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE UPDATED.
DURING THE INSPECTION OF THE RETURNED LONER INSTRUMENTS FROM THE HOSPITAL, IT WAS FOUND THAT THE TIP OF EXTRACTOR BROKE. ACCORDING TO SALES REP, THE PROBLEM WAS NOT SEEN DURING THE OPERATION.
DURING THE INSPECTION OF THE RETURNED LONER INSTRUMENTS FROM THE HOSPITAL, IT WAS FOUND THAT THE TIP OF EXTRACTOR BROKE. ACCORDING TO SALES REP, THE PROBLEM WAS NOT SEEN DURING THE OPERATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 544635 | EXTRACTOR ASNIS III FOR SCREWS:Ø4.0MM AO FITTING | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | STRYKER GMBH | H18318 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |