FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4062074 · Received September 4, 2014

Report

Report Number
2032227-2014-17191
Event Type
Malfunction
Date Received
September 4, 2014
Date of Event
August 5, 2014
Report Date
August 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY ALARM AND DISPLACEMENT TEST. NO MOTOR ERROR ALARM NOTED. THE MOTOR PASSED MOTOR TEST. MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, BROKEN BELT CLIP SLOT, AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED A MOTOR ERROR ALARM ON THE INSULIN PUMP. THE BLOOD GLUCOSE READING WAS 168 MG/DL. HE STATED THAT THE INSULIN PUMP HAD NOT BEEN EXPOSED TO A STRONG MAGNETIC FIELD OR MRI. CUSTOMER STATED THAT HE WAS ABLE TO REWIND THE INSULIN PUMP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
541674 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAB

Patients

Seq Age Sex Outcome Treatment
1 53 YR