FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4062069 · Received September 4, 2014

Report

Report Number
2032227-2014-16306
Event Type
Malfunction
Date Received
September 4, 2014
Date of Event
August 5, 2014
Report Date
August 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

NO UNEXPECTED EXCESSIVE NO DELIVERY ALARMS WERE NOTED. THE INSULIN PUMP PASSED THE DISPLACEMENT, PRIME AND EXCESSIVE NO DELIVERY ALARM TESTS. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, SCRATCHED RESERVOIR TUBE WINDOW AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT EVERY TIME HE INSERTS THE INFUSION SET ON HIS LEFT SIDE THE INSULIN PUMP ALARMS "NO DELIVERY." CUSTOMER STATED HE HAS USED THE BACK OF HIS THIGH AND THE DEVICE STILL ALARM "NO DELIVERY." CUSTOMER STATES THE DEVICE ONLY ALARMS NO DELIVERY DURING BOLUS DELIVERY. CUSTOMER DECLINED TROUBLESHOOTING AND REQUESTED A NEW PUMP. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
542923 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAB

Patients

Seq Age Sex Outcome Treatment
1 21 YR