FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4061449
·
Received September 4, 2014
Report
- Report Number
- 3004753838-2014-28346
- Event Type
- Malfunction
- Date Received
- September 4, 2014
- Date of Event
- August 10, 2014
- Report Date
- August 11, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(4). DEXCOM REVIEWED RECEIVER DATA ON (B)(6) 2014. THERE WERE INDICATIONS OF GAPS IN TRANSMISSION.
Additional Manufacturer Narrative · 1
(B)(4). SUBSEQUENT TO THE INITIAL MDR (B)(4), USED WITH THE COMPLAINT TRANSMITTER DEVICE, WAS RETURNED ON (B)(4) 2014. THE RETURNED COMPLAINT RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. EVALUATION OF THE RETURNED RECEIVER AND REVIEW OF THE DOWNLOADED DATA LOG CONFIRMED THE REPORTED EVENT OF INTERMITTENT ANTENNA ICON.
Description of Event or Problem · 1
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT AN OUT OF RANGE SIGNAL ON (B)(6) 2014. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 542094 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |