FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4061208 · Received September 4, 2014

Report

Report Number
2032227-2014-16778
Event Type
Malfunction
Date Received
September 4, 2014
Date of Event
August 5, 2014
Report Date
August 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE AND FLUSH DRIVE SUPPORT DISK. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A MOTOR ERROR ALARM ON THE INSULIN PUMP DURING PRIMING. CUSTOMER DOES NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ERROR ALARM. BLOOD GLUCOSE LEVEL AT THE TIME OF THE CALL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
540782 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722NAL A6722NALJ

Patients

Seq Age Sex Outcome Treatment
1 25 YR