FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 4060483 · Received September 4, 2014

Report

Report Number
2024168-2014-05673
Event Type
Malfunction
Date Received
September 4, 2014
Date of Event
July 21, 2014
Report Date
July 23, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K103110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED BALLOON RUPTURE WAS CONFIRMED. BASED ON VISUAL, DIMENSIONAL, FUNCTIONAL, AND SCANNING ELECTRON MICROSCOPY (SEM) IMAGING ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS FOR BALLOON RUPTURE REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 95% STENOSED, MILDLY CALCIFIED LESION IN THE HEAVILY TORTUOUS LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE 2.5X15MM VOYAGER NC BALLOON DILATATION CATHETER (BDC) BALLOON RUPTURED AT 18 ATMOSPHERES (ATM) AND THE BDC WAS REMOVED WITHOUT REPORTED ISSUE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER VOYAGER NC WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
542208 VOYAGER NC CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 3013161

Patients

Seq Age Sex Outcome Treatment
1