PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4060397
·
Received September 4, 2014
Report
- Report Number
- 1031452-2014-08126
- Event Type
- Malfunction
- Date Received
- September 4, 2014
- Report Date
- August 7, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER INDEPENDENT REPAIR CENTER STATEMENT, THE UNIT HAD LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS THE SIEVE BEDS LEAKING. ADDITIONAL MALFUNCTIONS WERE TIE WRAP INSTALLED, CLAMP INSTALLED, 4 WAY VALVE LEAKING AND POPPET VALVE WAS LEAKING.