FDA Adverse Event Malfunction Summary report: N

NEUROTHERM DISPOSABLE GROUNDING PAD

MDR report key: 4059385 · Received August 15, 2014

Report

Report Number
3002953813-2014-00008
Event Type
Malfunction
Date Received
August 15, 2014
Date of Event
July 17, 2014
Report Date
July 18, 2014
Manufacturer
NEUROTHERM, INC.
Product Code
GXD
PMA / PMN Number
K052878
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE GROUNDING PAD UNDERWENT VISUAL AND PERFORMANCE TESTING UPON RETURN. THREE 1 MINUTE LESIONS AT 90 DEGREES CELSIUS WERE PERFORMED. EACH PROCEDURE HELD TEMPERATURE AND IMPEDANCE WITHIN SPECIFICATION. (B)(4).

Description of Event or Problem · 1

A PATIENT RECEIVED A BURN FROM A GROUNDING PAD ON THE MID SECTION OF THE BACK BETWEEN L1 AND L2. THE BURN WAS LOCATED ON TWO SIDES WHERE THE PAD CURVES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
491458 NEUROTHERM DISPOSABLE GROUNDING PAD GXD NEUROTHERM, INC. RF-DGP-S 20140121

Patients

Seq Age Sex Outcome Treatment
1 Other