PINNACLE MTL INS NEUT36IDX52OD
Report
- Report Number
- 1818910-2014-27242
- Event Type
- Injury
- Date Received
- September 4, 2014
- Date of Event
- August 20, 2014
- Report Date
- October 6, 2014
- Manufacturer
- DEPUY INTERNATIONAL LTD. 8010379
- Product Code
- KWA
- PMA / PMN Number
- PK003523
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. PER WI-3430, A REVIEW OF THE DEVICE HISTORY RECORDS FOR THE 1900996 LOT CODE IS NO LONGER REQUIRED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PATIENT WAS REVISED TO ADDRESS PAIN. UPON REVISION METALLOSIS WAS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 541193 | PINNACLE MTL INS NEUT36IDX52OD | HIP ACETABULAR INSERT/LINER | KWA | DEPUY INTERNATIONAL LTD. 8010379 | 1890690 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention |