FDA Adverse Event
Other
Summary report: N
ADVANTA BED
MDR report key: 405550
·
Received July 15, 2002
Report
- Report Number
- 1824206-2002-00124
- Event Type
- Other
- Date Received
- July 15, 2002
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
07/2002 CUSTOMER STATES THAT A PATIENT CLAIMED THE BED HAD AN UNINTENTIONAL MOVEMENT OF THE HILOW FUNCTION BOTH UP AND DOWN. NO INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | P1600 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |