I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2014-00046
- Event Type
- Malfunction
- Date Received
- September 3, 2014
- Date of Event
- August 15, 2014
- Report Date
- August 15, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE INVESTIGATION WAS COMPLETED ON 09/05/2014. RETURN AND RETAIN CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
(B)(4).
ON (B)(4) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.28 ON A 53 YEAR OLD MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.DATE TIME METHOD RESULT SAMPLE(B)(6) 2014 06:43 I-STAT 0.28 A(B)(6) 2014 07:03 I-STAT 0.00 A (REPEAT)(B)(6) 2014 08:00 I-STAT 0.00 C(B)(6) 2014 UNK I-STAT 0.00 C (REPEAT)(B)(6) 2014 06:35 LAB CPK:147 B (CPK 49-397)THERE WERE NO INJURIES REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 539391 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE | NA | P14136 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |