FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 4053429 · Received September 3, 2014

Report

Report Number
2245578-2014-00046
Event Type
Malfunction
Date Received
September 3, 2014
Date of Event
August 15, 2014
Report Date
August 15, 2014
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 09/05/2014. RETURN AND RETAIN CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(4) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.28 ON A 53 YEAR OLD MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.DATE TIME METHOD RESULT SAMPLE(B)(6) 2014 06:43 I-STAT 0.28 A(B)(6) 2014 07:03 I-STAT 0.00 A (REPEAT)(B)(6) 2014 08:00 I-STAT 0.00 C(B)(6) 2014 UNK I-STAT 0.00 C (REPEAT)(B)(6) 2014 06:35 LAB CPK:147 B (CPK 49-397)THERE WERE NO INJURIES REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
539391 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE NA P14136

Patients

Seq Age Sex Outcome Treatment
1 53 YR