FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4052669 · Received September 2, 2014

Report

Report Number
2531779-2014-24934
Event Type
Malfunction
Date Received
September 2, 2014
Report Date
August 12, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERED ON WITH NO AUDIBLE TONES PRESENT. THE PUMP WAS OPENED AND THE AUDIBLE TONE CIRCUIT CONTACTS WERE OUT OF ALIGNMENT. THE CONTACTS WERE REALIGNED AND THE PUMP WAS FOUND TO BOOT WITH FUNCTIONAL AUDIO TONES. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS. PRODUCT ANALYSIS EVALUATED THE PUMP AND FOUND THAT THE PUMP AUDIBLE TONE WAS NOT WORKING. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
534797 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 47 YR