PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4052600 · Received September 2, 2014

Report

Report Number
2032227-2014-15601
Event Type
Malfunction
Date Received
September 2, 2014
Date of Event
August 4, 2014
Report Date
August 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.(B)(4).

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW, A MISSING END CAP STICKER AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 390 MG/DL. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.