FDA Adverse Event Malfunction Summary report: N

NEOPUFF INFANT RESUSCITATOR

MDR report key: 4050550 · Received September 1, 2014

Report

Report Number
9611451-2014-00742
Event Type
Malfunction
Date Received
September 1, 2014
Report Date
August 4, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTL
PMA / PMN Number
K892885
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO OUR FISHER & PAYKEL HEALTHCARE (FPH) REGIONAL OFFICE IN (B)(4), WHERE IT WAS INSPECTED BY A TRAINED FPH TECHNICIAN. RESULTS: VISUAL INSPECTION OF THE NEOPUFF REVEALED THAT THE GAS INLET PORT WAS BROKEN. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT NUMBER 070817. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. IT IS MOST LIKELY THAT THE PHYSICAL DAMAGE TO THE GAS INLET PORT, CASING AND MANOMETER WAS CAUSED BY SOME SORT OF IMPACT. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF/PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE GAS INLET PORT OF A NEOPUFF INFANT RESUSCITATOR WAS BROKEN AND REQUESTED A PERFORMANCE CHECK. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
531622 NEOPUFF INFANT RESUSCITATOR BTL BTL FISHER & PAYKEL HEALTHCARE LTD RD900 070817

Patients

Seq Age Sex Outcome Treatment
1