FDA Adverse Event Malfunction Summary report: N

PRECISION®

MDR report key: 4050239 · Received September 1, 2014

Report

Report Number
3006630150-2014-01932
Event Type
Malfunction
Date Received
September 1, 2014
Date of Event
June 2, 2014
Report Date
June 9, 2014
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
030017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT NO FURTHER COURSE OF ACTION WILL BE TAKEN AT THIS TIME. IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.

Additional Manufacturer Narrative · 1

EXPIRATION DATE: NI.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT UNDERWENT AN IPG REPLACEMENT. DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT WAS REPORTEDLY DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WAS NOT RETURNED TO BSN AS IT WAS DISCARDED BY THE MEDICAL FACILITY.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IMPLANT. IT DEPLETED QUICKER THAN IT USED TO. A BSN REPRESENTATIVE ANALYZED THE IPG DATABASE AND CONFIRMED PREMATURE BATTERY DEPLETION. IPG REPLACEMENT HAS BEEN RECOMMENDED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IMPLANT. IT DEPLETED QUICKER THAN IT USED TO. A BSN REPRESENTATIVE ANALYZED THE IPG DATABASE AND CONFIRMED PREMATURE BATTERY DEPLETION. IPG REPLACEMENT HAS BEEN RECOMMENDED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IMPLANT. IT DEPLETED QUICKER THAN IT USED TO. A BSN REPRESENTATIVE ANALYZED THE IPG DATABASE AND CONFIRMED PREMATURE BATTERY DEPLETION. IPG REPLACEMENT HAS BEEN RECOMMENDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
531545 PRECISION® SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1110 NA

Patients

Seq Age Sex Outcome Treatment
1 43 YR