FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4050141 · Received August 30, 2014

Report

Report Number
2032227-2014-14551
Event Type
Malfunction
Date Received
August 30, 2014
Date of Event
July 31, 2014
Report Date
July 31, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL OF THE BUTTONS FUNCTIONED PROPERLY DURING TESTING. HOWEVER, MOISTURE DAMAGE WAS FOUND ON THE KEYPAD TRACES DURING A VISUAL INSPECTION. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A CRACKED CASE AT A DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, AND A CRACKED BELT CLIP SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER HAD A BUTTON ERROR ALARM. THE CUSTOMER DID NOT EXPOSE THEIR INSULIN PUMP TO MOISTURE. NO ADDITIONAL INFORMATION PROVIDED.