PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-14551
- Event Type
- Malfunction
- Date Received
- August 30, 2014
- Date of Event
- July 31, 2014
- Report Date
- July 31, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER
Narratives
ALL OF THE BUTTONS FUNCTIONED PROPERLY DURING TESTING. HOWEVER, MOISTURE DAMAGE WAS FOUND ON THE KEYPAD TRACES DURING A VISUAL INSPECTION. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A CRACKED CASE AT A DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, AND A CRACKED BELT CLIP SLOT.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED THE CUSTOMER HAD A BUTTON ERROR ALARM. THE CUSTOMER DID NOT EXPOSE THEIR INSULIN PUMP TO MOISTURE. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 531405 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722WWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |