PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-14650
- Event Type
- Malfunction
- Date Received
- August 30, 2014
- Date of Event
- July 31, 2014
- Report Date
- July 31, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER
Narratives
(B)(4).THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM OR RAMPING NUMBERS NOTED. THE DEVICE WAS RECEIVED WITH CRACKED CASE ON THE DISPLAY WINDOW CORNERS, MINOR SCRATCHED LCD WINDOW, CRACKED RESERVOIR TUBE LIP, AND CRACKED BATTERY TUBE THREADS.
IT WAS REPORTED THAT THE CUSTOMER HAD A BUTTON ERROR ON THE INSULIN PUMP. IT WAS ALSO REPORTED THAT THE CASE OF THE INSULIN PUMP WAS CRACKED AS WELL AS THE RIGHT SIDE OF THE DISPLAY. INSULIN PUMP WILL NEED TO BE REPLACED. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 531079 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754WWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |