FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4050098 · Received August 30, 2014

Report

Report Number
2032227-2014-14575
Event Type
Malfunction
Date Received
August 30, 2014
Date of Event
July 31, 2014
Report Date
July 31, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. CRACKS TO THE LCD WINDOW, CASE, RESERVOIR TUBE LIP, BELT CLIP SLOT, BATTERY TUBE THREADS, A STAINED ADDRESS/SERIAL NUMBER LABEL AND A STAINED END CAP STICKER ALSO NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE BUTTONS ON THE INSULIN PUMP HAVE INTERMITTENT RESPONSE. THE DEVICE WAS NOT DROPPED OR EXPOSED TO MOISTURE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER DID NOT KNOW CURRENT BLOOD GLUCOSE READING. NO FURTHER INFORMATION REPORTED.