PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4050094 · Received August 30, 2014

Report

Report Number
2032227-2014-14626
Event Type
Malfunction
Date Received
August 30, 2014
Date of Event
July 31, 2014
Report Date
July 31, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE AND CRACKED RESERVOIR TUBE WINDOW WERE NOTED DURING THE VISUAL INSPECTION. NO BUTTON ERROR ALARMS WERE NOTED. THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE WAS NOTED IN THE BATTERY TUBE OR BATTERY CAP. THE INSULIN PUMP HAD MOISTURE DAMAGE ON THE DISPLAY BOARD AND MOTOR ASSEMBLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD A BUTTON ERROR ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 130 MG/DL. IT WAS FOUND THAT THERE WAS A POSSIBLE LEAK IN THE BATTERY COMPARTMENT. INSULIN PUMP WILL NEED TO BE REPLACED. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
531015 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 39 YR