XPER IM PHYSIOMONITORING 5 HOST
Report
- Report Number
- 1039368-2014-00002
- Event Type
- Malfunction
- Date Received
- June 6, 2014
- Manufacturer
- WITT BIOMEDICAL CORP, A WHOLLY OWNED SUBSIDIARY OF PHILIPS HOLDING USA, INC.
- Product Code
- DQK
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AS REPORTED ON (B)(6) 2013 BY (B)(6), SYSTEM ADMINISTRATOR FOR (B)(6) THEY WERE HAVING AN ISSUE WITH HAVING INACCURATE HEMODYNAMIC DATA. PHILIPS SUPPORT PERSONNEL FOUND THE ISSUE THEY WERE HAVING WAS WITH THE HEMODYNAMIC CONDITION LIST. IF THERE IS A SPACE AT THE END OF THE CONDITION THE CONDITION WILL DUPLICATE THEN RENAME THE SECOND CONDITION TO THE FIRST CONDITION IN THE LIST. PHILIPS SUPPORT PERSONNEL FOUND THE ISSUE THE ACCOUNT WAS HAVING WAS WITH THE HEMODYNAMIC CONDITION LIST. IF THERE IS A SPACE AT THE END OF THE CONDITION, THE CONDITION WILL DUPLICATE AND THEN RENAME THE SECOND CONDITION TO THE FIRST CONDITION IN THE LIST. PHILIPS SUPPORT PERSONNEL WERE ABLE TO ASSIST THE ACCOUNT TO TROUBLESHOOT THE REPORTED ISSUE AND RESOLVE IT BY REMOVING THE SPACES AT THE END OF THE CONDITION NAME. THE ACCOUNT HAS NOT CONTACTED PHILIPS WITH ANY FURTHER ISSUES WITH THE CONDITIONS LIST AFTER CUSTOMER ASSISTANCE WAS PROVIDED. PER THE REVIEW OF THE XPER INFORMATION MANAGEMENT SYSTEM PM 5 INSTRUCTIONS FOR USE RELEASE 4 (PART NUMBER PHILIPS HEALTHCARE 9896 060 00014 IPC P/N NMANXIM-ENG XPER IM PM5 MANUAL) THE CONDITION LIST IS CUSTOMIZABLE BY THE ACCOUNT AND IS A DATA ENTRY FIELD WHICH ALLOWS THE USER TO ADD SPACES. CONCLUSION: AS THIS CONDITION LIST IS CUSTOMIZABLE BY THE CUSTOMER(S) TO ADDRESS THE DIFFERENT NEEDS AT DIFFERENT ACCOUNTS, ENTRY ERRORS CAN OCCUR DURING THE GENERATION OF THAT LIST. CUSTOMER SERVICE PERSONNEL IS AVAILABLE TO ASSIST THE CUSTOMER(S) IN GENERATING THOSE LISTS; IN ADDITION PHILIPS HEALTHCARE PROVIDES USER TRAINING TO ACCOUNT REPRESENTATIVES AS NEEDED. THE REPORTED DEVICE REMAINS AT THE CUSTOMER SITE AND REMAINS IN CLINICAL USE AND IS FULLY FUNCTIONAL.
REFERENCE U/F (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333310 | XPER IM PHYSIOMONITORING 5 HOST | PHYSIO-MONITORING SYSTEM | DQK | WITT BIOMEDICAL CORP, A WHOLLY OWNED SUBSIDIARY OF PHILIPS HOLDING USA, INC. | 4522 300 34961 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |