FRESENIUS PD +IQ CARD CYCLER (NEWTON IQ)
Report
- Report Number
- 2937457-2014-02138
- Event Type
- Injury
- Date Received
- August 15, 2014
- Date of Event
- April 4, 2012
- Report Date
- April 4, 2012
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- FKX
- PMA / PMN Number
- K002892
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
A PRODUCT INVESTIGATION WAS PERFORMED FOR THIS DEVICE. A VISUAL INSPECTION WAS PERFORMED. THERE WERE NO DISCERNIBLE PHYSICAL DAMAGE TO THE NEWTON CYCLER. A STIMULATED TREATMENT WAS CONCLUDED WITH NO ALARMS OR ISSUES. THE REPORTED SYMPTOMS WERE NOT CONFIRMED. ALL FUNCTIONAL TESTS PASSED. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED. THERE WERE NO UNEXPECTED VARIANCES, ADJUSTMENTS OR COMMENTS NOTED IN THE DEVICE RECORDS. A SEARCH OF THE COMPLAINT TRACKING SYSTEM LISTED NO OTHER COMPLAINTS FOR THIS DEVICE WITH THE REPORTED SYMPTOMS. A REVIEW WAS PERFORMED BY THE POST MARKET CLINICAL DEPT. PER THE PHARMACOVIGILANCE ASSOCIATE MED DIR, IT IS UNLIKELY THAT THE CYCLER FILLED THE PATIENT WITH 7000 ML OF SOLUTION. THE TOTAL TREATMENT VOLUME OF 12000 ML AND TOTAL SOLUTION VOLUME IN BAGS: 15000 ML. IT IS HIGHLY UNLIKELY THAT THE CYCLER INFUSED THE PATIENT WITH THE LARGE VOLUME AND STILL HAD ENOUGH SOLUTION TO COMPLETE TREATMENT. THERE WAS NO REPORTED INJURY. THIS IS A RETROSPECTIVE MDR BEING SUBMITTED BY FRESENIUS MED CARE AS A REMEDIATION ACTION COMMITTED TO FDA REGARDING ALL POTENTIAL ADVERSE EVENTS RECEIVED BETWEEN 21 MARCH 2011 AND 11 MARCH 2013.
A PERITONEAL DIALYSIS PATIENT CALLED FRESENIUS TECHNICAL SUPPORT REGARDING ALARMS IN DRAIN 6, EMPTYING ALARM AND WEIGHT BAG EMPTY. THE PATIENT STATED THAT THE CYCLER HAS BEEN ALARMING ALL NIGHT. TECHNICAL SUPPORT REVIEWED THE DATA SHEET: FILL VOLUMES WERE 2000 ML. DRAIN 1 = 1760 ML, DRAIN 2 = 1760 ML, DRAIN 3 = 7160 ML, DRAIN 4 = 2250 ML, DRAIN 5 = 1910 ML, AND DRAIN 6 = 2310 ML. THE PATIENT REPORTED FEELING FULL, BUT NOT 7 L FULL. THE PATIENT COMPLETED THE FULL TREATMENT WITH 6 FILLS AND 6 DRAINS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 491147 | FRESENIUS PD +IQ CARD CYCLER (NEWTON IQ) | FKX | FRESENIUS MEDICAL CARE NORTH AMERICA | NEWTON | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |