RESERVOIR 3ML
Report
- Report Number
- 3004209178-2014-99704
- Event Type
- Malfunction
- Date Received
- August 27, 2014
- Date of Event
- July 27, 2014
- Report Date
- July 28, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
INSPECTED RESERVOIR, SNAP-CAP AND SEPTUM FOR ANOMALIES NONE WERE FOUND. PERFORMED OCCLUSION TEST PER SPECIFICATIONS, RESERVOIR FILLED AND CONNECTED TO NEW INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR INTO INSULIN PUMP AND PERFORMED INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR TO INSULIN PUMP AND PERFORMED CATHETER TIP. RESERVOIR NOT OCCLUDED.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING MOTOR ERROR WHILE ATTEMPTING TO PRIME. DURING TROUBLESHOOTING, CUSTOMER STATED THAT THEY DO NOT USE THE SENSOR FEATURE AND WAS ABLE TO REWIND THE INSULIN PUMP. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LEAD TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 112 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 519396 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A | H9156900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |