ALLIGATOR GRASPER
Report
- Report Number
- 3007123990-2014-00002
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Manufacturer
- MINI LAP TECHNOLOGIES INC.
- Product Code
- KOG
- PMA / PMN Number
- K070686
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
AS OF THIS DATE THE INSTRUMENT HAS NOT BEEN MADE AVAILABLE FOR MINI-LAP TECHNOLOGIES TO EVALUATE; THEREFORE, THE REPORTED FAILURE CANNOT BE CONFIRMED. A REVIEW OF THE DEVICE'S LOT HISTORY FOLDER WAS CONDUCTED AND NO ABNORMALITIES WERE NOTED AND SHOWED NO PREVIOUS RELATED NONCONFORMITIES. TO VERIFY THE STRENGTH AND RELIABILITY OF THE JAWS, MINILAP TECHNOLOGIES PERFORMED A SIMULATED INSTRUMENT UTILIZING AN INSTRON MACHINE TO EVALUATE HOW MUCH FORCE IT WOULD TAKE TO BREAK THE JAWS. THE RESULTS SHOWED THAT DURING A STRAIGHT PULL THE JAWS YIELDED AT AN AVERAGE OF 43.89 LBS PER LEG. ANOTHER TEST WAS PERFORMED ON 10 INSTRUMENTS BY BENDING EACH JAW 180 DEGREES FROM THEIR PRESENT POSITIONS TO DETERMINE HOW MANY BENDING CYCLES OF THE INDIVIDUAL JAW LEGS WOULD CAUSE THE DEVICE TO FAIL. THE TEN INSTRUMENTS (20) JAWS AVERAGED 14 BENDS PRIOR TO DEFORMATION (CRACKING, BREAKING, TWISTING). A CORRESPONDENCE WAS SENT TO OUR DISTRIBUTOR INSTRUCTING THEM TO CONTACT ALL THEIR SALES REPS TO INSTRUCT THE END USER TO BE COGNITIVE OF MANIPULATION AND TISSUE PLACED IN THE JAWS. A REVIEW OF THE INSTRUCTION FOR USE BOOKLET STATES UNDER WARNINGS AND PRECAUTIONS ITEM # 7, 8 TO AVOID DAMAGE TO THE INSTRUMENT TIPS (JAWS). ADDITIONAL TESTING OF THE JAWS WAS CONDUCTED UNDER PROTOCOLS TR-0066 JAW FORCE TEST AND TR-0069 JAW RELIABILITY TEST, VERIFYING STRENGTH AND RELIABILITY (AVG 63 LBS AND 185 LBS AND RELIABILITY PASSED THREE TIMES). PROBABLE ROOT CAUSES FOR THE REPORTED FAILURE INVOLVING THIS DEVICE APPEARS RELATED TO MISUSE OR OVEREXERTION BY THE USER WITH EXCESSIVE TISSUE GRASPED IN THE JAWS. WE ARE UNABLE TO OBTAIN FURTHER DETAILS FROM THE USER. THE COMPLAINT WILL BE CLOSED WITHOUT A DETAILED INVESTIGATION REPORT BASED ON PROBABLE ROOT CAUSE.
DURING A LAPAROSCOPIC CHOLECYSTECTOMY AN ALLIGATOR GRASPER TIP BROKE INTO THE PATIENT AND TIP RETRIEVED. THERE WAS NO HARM TO THE PATIENT. THERE WERE NO VISUAL DEFECTS TO THE GRASPER PRIOR TO USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379964 | ALLIGATOR GRASPER | KOG | MINI LAP TECHNOLOGIES INC. | 0255000001 | ML-000193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |