PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-11281
- Event Type
- Malfunction
- Date Received
- August 27, 2014
- Date of Event
- July 28, 2014
- Report Date
- August 30, 2016
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
ALL BUTTONS FUNCTIONED PROPERLY. HOWEVER, FOUND CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS DURING TESTING NOTED. THE PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, A CRACKED DISPLAY WINDOW, AND MINOR SCRATCHES ON THE DISPLAY WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER REVERTED TO THEIR OLD INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 521013 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754LWWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |