FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4040702 · Received August 27, 2014

Report

Report Number
3004209178-2014-98831
Event Type
Malfunction
Date Received
August 27, 2014
Date of Event
July 28, 2014
Report Date
July 28, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. TROUBLESHOOTING WAS DONE. CUSTOMER WAS MISSING PIECE ON THE BATTERY COMPARTMENT AND BELT CLIP NOT HOLDING THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE AT THAT TIME WAS 133 MG/DL. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
519988 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 31 YR