FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4040619 · Received August 27, 2014

Report

Report Number
2032227-2014-11223
Event Type
Malfunction
Date Received
August 27, 2014
Date of Event
July 25, 2014
Report Date
July 28, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA, BROKEN RESERVOIR TUBE LIP AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 217 MG/DL. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
522373 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722RNAP

Patients

Seq Age Sex Outcome Treatment
1 68 YR