FDA Adverse Event
Malfunction
Summary report: N
PAKAUTO ASSAY
MDR report key: 4036656
·
Received August 15, 2014
Report
- Report Number
- 2183608-2014-00009
- Event Type
- Malfunction
- Date Received
- August 15, 2014
- Date of Event
- April 25, 2013
- Report Date
- April 25, 2013
- Manufacturer
- IMMOCUR GTI DIAGNOSTICS, INC.
- Product Code
- MZI
- PMA / PMN Number
- BK960059
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NA
Description of Event or Problem · 1
A CUSTOMER COMPLAINT WAS RECEIVED (CMP-76), WHERE CUSTOMER WAS SEEING VERY LOW O.D. VALUES (-0.01 - 0.014) WITH THEIR IN-HOUSE NEGATIVE ELUATE CONTROL (NCP), WHEN USING PAKAUTO ASSAY LOT 300655. THIS LEAD TO LOW ASSAY CUT-OFF VALUE (0.020 - 0.028), WHICH RESULTED IN ALMOST EVERY PATIENT ON THE PLAT HAVING A POSITIVE RESULT WITH ALMOST ALL GLYCOPROTEINS (GPS). CUSTOMER INFORMED THA THEY OBTAINED LOW OD VALUES WITH BOTH THEIR IN-HOUSE NCP AND KIT NCP, AND OCCURRENCE OF THE LOW NCP OD VALUES SEEMED SPORADIC. CUSTOMER'S QC DATA TRACKING OVER THE YEARS SHOW THAT THE NCP OD VALUE TYPICALLY RUNS BETWEEN 0.030-0.068.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 490487 | PAKAUTO ASSAY | PAKAUTO ASSAY | MZI | IMMOCUR GTI DIAGNOSTICS, INC. | NA | 300655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |