FDA Adverse Event Malfunction Summary report: N

PAKAUTO ASSAY

MDR report key: 4036656 · Received August 15, 2014

Report

Report Number
2183608-2014-00009
Event Type
Malfunction
Date Received
August 15, 2014
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
IMMOCUR GTI DIAGNOSTICS, INC.
Product Code
MZI
PMA / PMN Number
BK960059
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NA

Description of Event or Problem · 1

A CUSTOMER COMPLAINT WAS RECEIVED (CMP-76), WHERE CUSTOMER WAS SEEING VERY LOW O.D. VALUES (-0.01 - 0.014) WITH THEIR IN-HOUSE NEGATIVE ELUATE CONTROL (NCP), WHEN USING PAKAUTO ASSAY LOT 300655. THIS LEAD TO LOW ASSAY CUT-OFF VALUE (0.020 - 0.028), WHICH RESULTED IN ALMOST EVERY PATIENT ON THE PLAT HAVING A POSITIVE RESULT WITH ALMOST ALL GLYCOPROTEINS (GPS). CUSTOMER INFORMED THA THEY OBTAINED LOW OD VALUES WITH BOTH THEIR IN-HOUSE NCP AND KIT NCP, AND OCCURRENCE OF THE LOW NCP OD VALUES SEEMED SPORADIC. CUSTOMER'S QC DATA TRACKING OVER THE YEARS SHOW THAT THE NCP OD VALUE TYPICALLY RUNS BETWEEN 0.030-0.068.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
490487 PAKAUTO ASSAY PAKAUTO ASSAY MZI IMMOCUR GTI DIAGNOSTICS, INC. NA 300655

Patients

Seq Age Sex Outcome Treatment
1 NA