FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4033678 · Received August 25, 2014

Report

Report Number
2032227-2014-10431
Event Type
Malfunction
Date Received
August 25, 2014
Date of Event
July 24, 2014
Report Date
July 24, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD NO MOISTURE DAMAGE FOUND ON THE ELECTRONICS, MOTOR, VIBRATOR, AND BATTERY TUBE ASSEMBLY DURING VISUAL INSPECTION. INSULIN PUMP RECEIVED WITH NO BUTTON RESPONSE DUE TO CORROSION ON KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED DURING TESTING. INSULIN PUMP HAD A CRACKED CASE ON THE LCD DISPLAY WINDOW CORNERS, CRACKS ON THE BATTERY TUBE THREADS, CRACKS ON THE RESERVOIR TUBE LIP AND CRACKS ON THE RESERVOIR TUBE.

Description of Event or Problem · 1

IT IS REPORTED THAT THE CUSTOMER HAD HIGH BLOOD GLUCOSE LEVELS OF 144 MG/DL. THE CUSTOMER REPORTED A BUTTON ERROR ALARM FROM THE INSULIN PUMP. THE CUSTOMER REPORTED WEARING THE INSULIN PUMP INSIDE THE BRA AND SHE WAS SWEATING. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
512929 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 70 YR