FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 4033668 · Received August 25, 2014

Report

Report Number
2531779-2014-24270
Event Type
Malfunction
Date Received
August 25, 2014
Report Date
August 11, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE BATTERY CAP WAS NOT RETURNED AND A TEST CAP WAS USED IN THE EVALUATION. THE PUMP POWERED UP TO A DIM AND DISCOLORED DISPLAY WITH AUDIBLE AND VIBRATORY FEATURES. THE BATTERY COMPARTMENT WAS CRACKED AT THE THREADS AND EVIDENCE OF MOISTURE INTRUSION WAS FOUND INSIDE THE COMPARTMENT. A LEAK TEST SHOWED A LEAK WHERE THE CRACK WAS LOCATED. PUMP CASING WAS OPENED AND NO EVIDENCE OF MOISTURE INGRESS WAS FOUND INSIDE THE PUMP. UNRELATED TO THE DISPLAY AND CASING ISSUE, THE BOLUS BUTTON COVER AND THE GRIP PAD WERE DAMAGED/PEELING. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED, MOISTURE INTRUSION WAS EVIDENT IN THE COMPARTMENT, AND THE DISPLAY WAS DIM/DISCOLORED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514379 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1