FDA Adverse Event Malfunction Summary report: N

COLIBRI

MDR report key: 4033645 · Received August 25, 2014

Report

Report Number
8030965-2014-00848
Event Type
Malfunction
Date Received
August 25, 2014
Report Date
January 29, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(6). DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE CONTROL UNIT DID NOT FUNCTION PROPERLY. THIS WAS ATTRIBUTED TO NORMAL WEAR. INSPECTION AND MAINTENANCE WERE PERFORMED, MOTOR AND CONTROL UNIT WERE REPLACED, AND THE DEVICE WAS RETURNED TO THE CUSTOMER ON (B)(6) 2013. PLACEHOLDER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE MACHINE DID NOT WORK. THE BATTERY AND THE BATTERY CASE GOT VERY HOT. THIS IS REPORT 1 OF 1 FOR COMPLAINT # (B)(4).