FDA Adverse Event Malfunction Summary report: N

COLIBRI

MDR report key: 4033644 · Received August 25, 2014

Report

Report Number
8030965-2014-00674
Event Type
Malfunction
Date Received
August 25, 2014
Report Date
February 27, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO (B)(4) AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(6). 510K#: DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE TRIGGER DID NOT FUNCTION PROPERLY AND THE MOTOR AND CONTROL UNIT DREW TOO MUCH POWER. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS RETURNED TO THE CUSTOMER ON (B)(4) 2013. PLACEHOLDER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR SERVICE. THIS IS REPORT 1 OF 1 FOR COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514910 COLIBRI HWE SYNTHES GMBH 2200

Patients

Seq Age Sex Outcome Treatment
1