PARADIGM INSULIN INFUSION PUMP
MDR report key: 4033541
·
Received August 25, 2014
Report
- Report Number
- 3004209178-2014-96942
- Event Type
- Malfunction
- Date Received
- August 25, 2014
- Date of Event
- July 24, 2014
- Report Date
- July 24, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED FOR A BUTTON ERROR AND HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. IT WAS NOTED THAT THE ALARM PERSISTED EVEN AFTER A BATTERY CHANGE. NO FURTHER INFORMATION REPORTED.