FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4033254 · Received August 25, 2014

Report

Report Number
2032227-2014-10344
Event Type
Malfunction
Date Received
August 25, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY ALARM TESTS. NO UNEXPECTED MOTOR ERROR ALARM WAS NOTED. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. THE CUSTOMER'S BLOOD GLUCOSE WAS 197 MG/DL. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
512116 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 11 YR