FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4033222 · Received August 25, 2014

Report

Report Number
3004209178-2014-97664
Event Type
Malfunction
Date Received
August 25, 2014
Date of Event
July 24, 2014
Report Date
July 24, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED THE DISPLACEMENT, BASIC OCCLUSION, OCCLUSION, EXCESSIVE NO DELIVERY, PRIME AND MOTOR TESTS. NO MOTOR ERROR ALARM WAS OBSERVED DURING TESTING. CRACKS TO THE DISPLAY WINDOW AND RESERVOIR TUBE LIP NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. NO FURTHER INFORMATION PROVIDED.