FDA Adverse Event Malfunction Summary report: N

5.0CM SHORT ATTACHMENT

MDR report key: 4033016 · Received August 25, 2014

Report

Report Number
1045834-2014-12643
Event Type
Malfunction
Date Received
August 25, 2014
Report Date
August 5, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

\THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. (B)(4) EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE MET ALL MANUFACTURE'S SPECIFICATIONS. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED. AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EAR ATRESIA SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS OVERHEATING. THE REPORTER ALSO STATED THAT THE BURR SEEMED LOOSE. THERE WAS A ONE TO TWO MINUTES DELAY TO THE SURGICAL PROCEDURE TO SET-UP THE SPARE IDENTICAL DEVICE. THE REPORTER STATED THAT THE SURGERY WAS COMPLETED SUCCESSFULLY USING THE SPARE IDENTICAL DEVICE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514330 5.0CM SHORT ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 CUTTING BURR DEVICE