PARADIGM INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-97329
- Event Type
- Malfunction
- Date Received
- August 25, 2014
- Date of Event
- July 23, 2014
- Report Date
- July 23, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, AND EXCESSIVE NO DELIVERY ALARM TESTS. NO UNEXPECTED MOTOR ERROR ALARM WAS NOTED. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND A STAINED END CAP STICKER.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR THREE TIMES. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 512741 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-715WWL |