FDA Adverse Event Injury Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 4031751 · Received August 22, 2014

Report

Report Number
2938836-2014-14766
Event Type
Injury
Date Received
August 22, 2014
Date of Event
June 5, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE EMERGENCY ROOM FOR INAPPROPRIATE SHOCKS DUE TO NOISE. NOISE WAS REPRODUCIBLE WITH ARM MOVEMENTS. CLAVICULAR CRUSH WAS SUSPECTED. THE LEAD WAS CAPPED AND REPLACED WITHOUT ANY COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509756 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7000/60 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention