PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4031738 · Received August 22, 2014

Report

Report Number
3004209178-2014-97188
Event Type
Injury
Date Received
August 22, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FUNCTIONED PROPERLY DURING THE REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY AND DISPLACEMENT TESTS. THE INSULIN PUMP WAS TESTED WITH WATER FILLED RESERVOIR. NO AIR BUBBLE WAS NOTED IN THE RESERVOIR. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT DISPLAY WINDOW CORNERS, A CRACKED RESERVOIR TUBE LIP AND A STAINED END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED HIS BLOOD GLUCOSE WAS HIGH AT 555 MG/DL, TREATED WITH MANUAL INJECTIONS. CUSTOMER'S SYMPTOM WAS EXCESSIVELY THIRSTY. TROUBLESHOOTING WAS PERFORMED. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE DEVICE. THE DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER DISCONNECTED AT THE QUICK RELEASE. NO LEAKS OR AIR BUBBLES NOTED. MANUAL PRIME WAS PERFORMED ON THE DEVICE AND INSULIN DID EXIT. CUSTOMER STATED HE SAW WHAT LOOKED LIKE BURNT MARKS ON THE INSIDE OF THE MEDTRONIC LOGO NEAR THE DRIVE SUPPORT CAP. CUSTOMER WAS UNAWARE OF HOW DAMAGE OCCURRED TO THE DEVICE BUT HE STATED IT LOOKED LIKE SOMETHING CAUSED THE BURNT MARKS FROM INSIDE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509354 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 10 YR Required Intervention