FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 4031651 · Received August 22, 2014

Report

Report Number
2955842-2014-05153
Event Type
Malfunction
Date Received
August 22, 2014
Date of Event
June 16, 2014
Report Date
July 28, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO CONFIRM THE REPORTED COMPLAINT OF THE INSTRUMENT STICKING. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE SYSTEM AND DRIVEN. THE GRIPS OPENED AND CLOSED PROPERLY AND THERE WERE NO WIRES, CABLES, OR OTHER FRAGMENTS STICKING OUT FROM THE INSTRUMENT. FAILURE ANALYSIS INVESTIGATION FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS INVESTIGATION, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTIONS WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE PROGRASP FORCEPS INSTRUMENT STICK TISSUE. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
508776 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-11 M11140220 340

Patients

Seq Age Sex Outcome Treatment
1 71 YR